Remote Control Commands
R&S
®
NRX
357
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:WIDTh
<value>
Effective if
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:MODE
USER
is set.
Sets the burst width.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 1.0
*RST:
0.01
Default unit: s
Manual operation:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:CCDF:THReshold
<value>
Sets the threshold for the complementary cumulative distribution function (CCDF).
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-290.0 to +110.0
*RST:
+0.0
Default unit: dBm
Manual operation:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:DIRection
<direction>
Defines how the forward power is determined.
Suffix:
<Measurement>
.
1 to 8
Measurement
Configuring Sensors
www.allice.de
Allice Messtechnik GmbH