Remote Control Commands
R&S
®
NRX
296
User Manual 1178.5566.02 ─ 07
13.6.2.2
Continuous Average Calculation Functions with Buffering
Parameter list
The following parameters are used.
●
<buffered_size>
Mandatory. Number of requested measured values.
Corresponds to
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
●
<expected_value>
Optional. Value that is expected for the measurement.
●
<resolution>
Optional. Limit up to which the measurement result should be free of noise.
Corresponds to
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
●
<source_list>
Mandatory. Primary and secondary sensor. The number for each sensor is prece-
ded by the character
@
. The entire expression is enclosed in parentheses.
Example: (
@
3),(
@
2)
Sensor C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
FETCh<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
READ<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
MEASure<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
CALCulate<Measurement>:MATH[:EXPRession]
"(SENS1)"
to
"(SENS4)"
.
Measured average power measured by one power sensor with buffering. The used
parameters are described in
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<buffered_size>
<expr>
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
Calculation Functions
www.allice.de
Allice Messtechnik GmbH