136
Diagnostic Command Reference
8
At the
diagmode >
prompt, proceed with the appropriate diagnostic tests.
Restoring Normal Mode
Follow this procedure if the SNC 4000 was placed in diagnostic mode and needs to be restored to normal
operation.
1
Connect the terminal to the service port, if it is not already connected.
2
If it is not already on, turn on the Scalar 24 and wait for it to finish booting.
3
From the service terminal, type the
normalBoot
command.
4
Wait for the Scalar 24 to reboot.
5
At the command prompt, continue with the MAP that sent you here.
Special Procedures
This section contains information on special procedures.
Health Check
The Health Check function provides a method to periodically determine the operational state of an SNC
4000 and its attached interfaces and devices.
Manual Health Check
Health Check is performed manually from the service terminal only if the SNC 4000 is in normal mode.
See
Perform Health Check
on page 44, for instructions.
Event Log Dump
The events that have been logged in the SNC 4000’s Event Log can be displayed on the service terminal
or viewed from the ADIC Management Console. For more information, see the
ADIC Management Console
User’s Guide
.
This information may be useful because you can see the normal events and any abnormal events that may
lead you to the source of an internal or external failure.
Each entry contains a sequence number, day, and time-stamp, the event code, event source (interface
name or system process), an index, and a description of the event.
If the Real Time Clock has been set, the time-stamp will be real time. Otherwise, the time stamp will be
relative to the boot-time.
The Event Log holds more than 2000 events. When the log becomes full, the oldest events are overwritten
by the most recent events.
You can choose to view only those events logged since the SNC 4000 was last booted or you can view a
longer list of events that have accumulated from previous boots.
You can choose to view events by severity level. Selecting
Warning
will display
Warning
and
Notice
events. Selecting
Information
will display
Information
,
Warning
and
Notice
events.
See the
loggerDump [number]
on page 92
or the
loggerDumpCurrent [level]
on page 93.
See also
Figure 8
on 37.
Содержание SNC 4000
Страница 1: ...www adic com ADICSNC 4000User sGuide ADVANCED DIGITAL INFORMATION CORPORATION...
Страница 10: ...4 SNC 4000 Introduction...
Страница 40: ...34 Removal and Replacement Procedures...
Страница 140: ...134 Service Port Command Reference...
Страница 147: ...SNC 4000 User s Guide 141 fcConnPortModeGet...
Страница 148: ...142 Diagnostic Command Reference...