Rotor-Gene Q MDx CE User Manual 02/2022
73
5.
Insert the assembly into the Rotor-Gene Q MDx chamber by clicking into place using the
locating pin on the rotor hub. To remove, simply push down on the rotor hub to release and
pull out.
6.
Close the lid and set up the run profile using the Rotor-Gene Q software.
5.2.3
Rotor-Disc setup
The Rotor-Disc 72 or Rotor-Disc 100 comprise 72 or 100 wells respectively in a one-piece disc
designed for high throughput. The Rotor-Disc 72 and Rotor-Disc 100 do not use caps. Instead, Rotor-
Disc Heat Sealing Film is applied to the top and heat sealed using a Rotor-Disc Heat Sealer. The
film prevents contamination by providing a strong, durable, and tamper-proof seal. Heat sealing
the Rotor-Disc is performed as described below.
Important
: Please read the Product Sheet supplied with the Rotor-Disc Heat Sealer before beginning
this procedure.
1.
Switch on the Rotor-Disc Heat Sealer using the switch located on the back at the left-hand side.
A red “Power” light illuminates. The Rotor-Disc Heat Sealer takes approximately 10 min to
reach operating temperature, when a green “Ready” light illuminates.
2.
Choose permanent or removable seal.
Note
: Once the Rotor-Disc Heat Sealer is ready, it is safe to leave it running constantly.
3.
Insert the Rotor-Disc into the Rotor-Disc Loading Block using the position one tab on the Rotor-
Disc and the tube guide holes on the Rotor-Disc Loading Block.