Rotor-Gene Q MDx CE User Manual 02/2022
24
Illustration of the optical system.
3.1.3
Available channels
Channel
Excitation (nm) Detection (nm) Examples of fluorophores detected
Blue
365 ± 20
460 ± 20
Marina Blue
®
, Edans Bothell Blue, Alexa Fluor
®
350,
AMCA-X, ATTO 390
Green
470 ± 10
510 ± 5
FAM
®
, SYBR
®
Green I, Fluorescein, EvaGreen
®
, Alexa
Fluor 488
Yellow
530 ± 5
557 ± 5
JOE™, VIC
®
, HEX™, TET™, CAL Fluor
®
Gold 540,
Yakima Yellow
®
Orange
585 ± 5
610 ± 5
ROX, CAL Fluor Red 610, Cy
®
3.5, Texas Red
®
, Alexa
Fluor 568
Red
625 ± 10
660 ± 10
Cy5, Quasar
®
670, LightCycler
®
Red640, Alexa Fluor
633
Crimson
680 ± 5
712 high pass Quasar 705, LightCycler Red705, Alexa Fluor 680
High resolution melt (HRM)
460 ± 20
510 ± 5
SYBR Green I, SYTO
®
9, LC Green
®
, LC Green Plus+,
EvaGreen
Note
: QIAGEN kits indicated for use with the Rotor-Gene Q MDx instruments are optimized with
respect to certain dye combinations. Please refer to the corresponding kit handbooks for more
information.