Rotor-Gene Q MDx CE User Manual 02/2022
19
WARNING
Damage to the instrument
When Rotor-Gene Q MDx is started immediately after delivery in cold climates,
mechanical parts can block.
Allow the instrument to acclimatize to room temperature for at least an hour
before turning the instrument on.
WARNING
Damage to the instrument
In case of breakdown caused by power failure, remove the power cord and
wait 10 min before attempting to manually open the lid.
WARNING
Risk of overheating
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
sides and rear of the Rotor-Gene Q MDx.
Slits and openings that ensure the ventilation of the Rotor-Gene Q MDx must not
be covered.
Heat hazard
WARNING
Hot surface
The Rotor-Gene Q MDx chamber can reach temperatures above 120°C
(248°F). Avoid touching it when it is hot.
WARNING
Hot surface
When pausing a run, the Rotor-Gene Q MDx will not be cooled completely to
room temperature. Exercise caution before handling the rotor or any tubes in
the instrument.