Chapter 4
| Monitoring
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
93
Troubleshooting Tips
Table 15
Troubleshooting Tips for EtCO2 Monitoring
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
ALARM APNEA
message
appears and waveform is solid
line at or near zero
No breath has been detected
for 30 seconds since last valid
breath
•
Check the patient.
FilterLine connection to device
is loose
•
Twist FilterLine connector
clockwise until tight and
firmly seated.
FilterLine set is disconnected
from patient or ETT
•
Check ventilation equipment
(if used) for leaks or
disconnected tubing.
CO2 FILTERLINE OFF
message
appears and waveform is "---"
FilterLine set disconnected or
not securely connected to
device
•
Connect FilterLine set to
device port.
•
Twist FilterLine connector
clockwise until tight and
firmly seated.
CO2 FILTERLINE PURGING
message appears and
waveform is "---"
FilterLine set is kinked or
clogged with fluid, or rapid
altitude change occurred
•
Disconnect and then
reconnect the FilterLine set.
•
Twist FilterLine connector
clockwise until tight and
firmly seated.
CO2 FILTERLINE BLOCKAGE
message appears and
waveform is "---"
The message appears
after 30 seconds of
unsuccessful purging
•
Disconnect and then
reconnect the FilterLine set.
•
Change the FilterLine set.
FilterLine set is kinked or
clogged
•
Twist FilterLine connector
clockwise until tight and
firmly seated..
CO2 INITIALIZING
message
appears and waveform is "---"
FilterLine set connected to
device while module is
initializing
•
None.
Defibrillation shock delivered
•
None. System resets
automatically
within 20 seconds.
AUTO ZEROING
message
appears and waveform is "---"
Module is performing self-
maintenance
•
None.
Defibrillation shock delivered
•
None. System resets
automatically
within 20 seconds.
EtCO
2
values are erratic
FilterLine connection to device
is loose
•
Twist FilterLine connector
clockwise until tight and
firmly seated.
A leak in the FilterLine set
•
Check for connection leaks
and line leaks to patient, and
correct, if necessary.
A mechanically ventilated
patient breathes spontaneously
or patient is talking
•
No action required.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......