Summary of Screen Messages
248
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MESSAGE
DESCRIPTION
OBTAINING DEVICE NAMES
Device is obtaining names of available
Bluetooth
-enabled
devices.
PACER FAULT
Internal error detected during pacing.
PACING IN PROGRESS
The requested action is not available because the device is
currently performing pacing.
PACING STOPPED
Pacing has stopped—for example, due to disconnection of
therapy electrodes.
PASSCODE INCORRECT -
TRY AGAIN
Incorrect passcode entered.
PAUSED
The pacing
PAUSE
button is pressed and held. Current pulses
are applied at reduced frequency while the MA and PPM
settings are maintained.
PUSH ANALYZE
Press
ANALYZE
to begin ECG analysis.
PUSH AND HOLD SHOCK
BUTTON!
The defibrillator is in Sync mode, fully charged, and ready to
provide therapy.
PUSH AND HOLD PADDLE
BUTTONS TO SHOCK!
The defibrillator is in Sync mode, fully charged, and ready to
provide therapy with hard paddles connected.
PUSH SHOCK BUTTON!
The defibrillator is fully charged and ready to provide therapy.
PX NOT ZEROED
Transducer is connected or reconnected and is not zeroed.
PX TRANSDUCER NOT DETECTED IP transducer is disconnected from the monitor/defibrillator.
PX ZERO FAILED
The device was unable to zero the pressure transducer.
PX ZEROED
Transducer successfully zeroed.
PX ZEROING
Monitor is establishing a zero reference.
RA LEADS OFF
ECG electrode "RA" is disconnected.
REPLACE BATTERY X
Power loss for the battery in well X is imminent.
SEARCHING FOR DEVICES
Device is attempting to identify available
Bluetooth
devices.
SELECT BIPHASIC ENERGY / XXX
J
ENERGY SELECT
was pressed on front panel or on standard
paddles.
SELF TEST FAILED
Device detected internal error; remove device from service.
SELF TEST FAILED.
TRANSMITTING...
Device detected internal error and is transmitting test results.
Remove device from service after transmission is complete.
SELF TEST IN PROGRESS
Device is performing a self test after turning on.
SELF TEST PASSED
Device passed internal test and is available for use.
SELF TEST PASSED.
TRANSMITTING...
Device passed internal test and is transmitting test results.
SHOCK ADVISED!
The defibrillator has analyzed the patient ECG rhythm and
detected a shockable ECG rhythm.
SPCO: POOR QUALITY SIGNAL
Device is not receiving sufficient input from sensor.
SPMET: POOR QUALITY SIGNAL
Device is not receiving sufficient input from sensor.
SPO2: CHECK SENSOR
The SpO
2
sensor connection to device or application to patient
needs checked.
SPO2: LOW PERFUSION
Patient has a weak pulse.
SPO2: NO SENSOR DETECTED
A sensor is disconnected from the monitor.
SPO2: POOR QUALITY SIGNAL
Device is not receiving sufficient input from sensor.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......