Chapter 5
| Therapy
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
123
Note:
If you do not press the (shock)
button within 60 seconds, or the
SPEED DIAL
is pressed to cancel charging, the
defibrillator disarms and the
DISARMING
message appears.
After a shock is delivered, the
START CPR
prompts occur. A countdown timer (min:sec
format) continues for the duration specified in
the
CPR TIME 1
setup option.
Note:
The CPR metronome automatically
provides audible compression “tocks” and
ventilation prompts or tones only during CPR
intervals at a ratio of 30:2. To silence the
metronome, press
CPR
. To un-silence the
metronome, press
CPR
again.
When the CPR countdown time ends, the
PUSH ANALYZE
prompts occur. These
prompts repeat every 20 seconds until you
press
ANALYZE
.
No Shock Advised
The following prompts occur if no shock is advised:
If the AED detects a nonshockable
rhythm, the
NO SHOCK ADVISED
prompts
occur. The defibrillator does not charge,
and no shock can be delivered.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......