Monitoring Invasive Pressure
102
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Failure to obtain PWP
Malposition of catheter tip
•
Reposition catheter.
Leak in balloon.
Ruptured balloon.
•
Replace catheter.
Progressive elevation of PWP
Overinflation
•
Inflate balloon in small
increments while watching
scope for confirmation of
wedging. Use only enough air
to wedge. Do not use more
than the volume
recommended by the
manufacturer.
Catheter migrated distally in
pulmonary artery
•
Reposition catheter.
For general troubleshooting tips, see General Troubleshooting Tips (on page 214).