Chapter 5
| Therapy
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
121
To perform automated external defibrillation:
1.
Verify that the patient is in cardiopulmonary arrest (unconscious, pulseless, not breathing
normally).
2.
Press
ON
.
3.
Prepare the patient for electrode placement (see Patient Skin Preparation (on page 117)).
The
CONNECT ELECTRODES
prompts
occur until the patient is connected to the
AED. If possible, place the patient on a hard
surface away from standing water.
4.
Connect the therapy electrodes to the therapy cable and confirm cable connection to the
defibrillator.
5.
Apply the therapy electrodes to the patient’s chest in the anterior-lateral position (see
Anterior-Lateral Placement (on page 117)).
The
PUSH ANALYZE
prompts occur when
the patient is properly connected to the
AED.
6.
Press
ANALYZE
to initiate the analysis. Stop CPR.
Possible Misinterpretation of Data
Do not move the AED during analysis. Moving the AED
during analysis may affect the ECG signal resulting in
an inappropriate
SHOCK
or
NO SHOCK ADVISED
decision. Do not touch the patient or the AED during
analysis.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......