Chapter 4
| Monitoring
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
79
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
SPMET: POOR QUALITY
SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
•
Check that sensor and cable
are connected properly.
•
Move sensor to a better
perfused site.
SPCO/SPMET: POOR QUALITY
SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
•
Check that sensor and cable
are connected properly.
•
Move sensor to a better
perfused site.
SPO2: SENSOR DOES NOT
SUPPORT SPCO OR SPMET
message appears
SpO
2
-only sensor used with
SpCO/SpMet capable device
•
None necessary, or use
Rainbow sensor to measure
SpCO or SpMet.
Note:
Most Rainbow sensor messages (SpO
2
, SpCO, and SpMet) are reported as
SPO2:
(MESSAGE)
. The
POOR QUALITY SIGNAL
message indicates the specific parameter affected.
For general troubleshooting tips, see General Troubleshooting Tips (on page 214).
Monitoring Noninvasive Blood Pressure
Intended Use
The LIFEPAK 15 noninvasive blood pressure (NIBP) monitor measures blood pressure (BP) using
the oscillometric measurement technique to determine systolic, diastolic, and mean arterial
pressures, and pulse rate. The measurement can be initiated manually or set to recur
automatically at predetermined intervals.
Blood pressure measurements determined using this device are equivalent to those obtained by
a trained observer using the cuff/stethoscope auscultation method, within the limits prescribed
by the American National Standard
Electronic or automated sphygmomanometers
(AAMI SP-10).
NIBP is a tool to be used in addition to patient assessment. Care should be taken to assess the
patient at all times; do not rely solely on the NIBP monitor.
Indications
Noninvasive blood pressure monitoring is intended for detection of hypertension or
hypotension and monitoring BP trends in patient conditions such as, but not limited to,
shock, acute dysrhythmia, or major fluid imbalance. NIBP monitoring is not indicated for
neonatal patients less than one month old.
Contraindications
None known.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......