Chapter 3
| Basic Orientation
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
33
Batteries
The LIFEPAK 15 monitor/defibrillator operates either on battery power using two Lithium-ion
batteries, or with auxiliary power using the AC Power Adapter or DC Power Adapter. Batteries
may be charged in the Station or Mobile Li-ion Battery Charger, the REDI-CHARGE™ Battery
Charger, or in the monitor/defibrillator if it is connected to auxiliary power.
Note:
Although the monitor/defibrillator can operate using auxiliary power with no batteries
installed, at least one battery should be installed at all times. If the monitor/defibrillator loses
power for more than 30 seconds, the device reverts to the user-configured default settings and
begins a new patient record.
IMPORTANT!
The LIFEPAK 15 monitor/defibrillator Lithium-ion batteries are not
interchangeable with batteries that are used in other LIFEPAK defibrillators.
Routinely inspect batteries for damage or leakage. Recycle or discard damaged or leaking
batteries.
Each battery has a fuel gauge that indicates the approximate charge level in the battery. Press
the gray button above the battery symbol to check the battery’s charge level prior to installing it
in the defibrillator. The four battery indicators shown here represent approximate charge—
greater than 70%, greater than 50%, greater than 25%, and 25% or less, respectively.
Figure 12
Battery Charge Indicators
Battery warning indicators are shown below. A single flashing LED indicates that the battery is
very low and needs to be charged. Any two or more flashing LEDs indicate that the battery is
faulty and should be returned to your authorized service personnel.
Figure 13
Battery Warning Indicators
Note:
Older or heavily used batteries lose charge capacity. If a battery fuel gauge indicates
fewer than four LEDs immediately after completing a charge cycle, the battery has reduced
capacity. If the battery fuel gauge shows two or fewer LEDs after the battery completes a charge
cycle, the battery should be replaced.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......