Chapter 4
| Monitoring
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
97
When the default labels P1 and P2 are used, the IP monitoring area displays systolic, diastolic,
and mean pressures. When ICP, LAP, or CVP labels are used, the IP monitoring area displays
mean pressure in large type. Systolic and diastolic pressures are not displayed.
Figure 28
IP Labels
FIGURE LEGEND
1
ART mean pressure
2
ART systolic pressure
3
ART diastolic pressure
4
CVP mean pressure
Because pressures can change in a short time, data should be checked regularly during vital
sign monitoring.
How IP Monitoring Works
IP monitoring involves the conversion of fluid pressure into an electrical signal. The conversion is
accomplished with a pressure transducer. The transducer is connected to a patient’s indwelling
pressure catheter using a special assembly of tubing, stopcocks, adapters, flush valves, and
fluids, commonly known as a flush system. The transducer translates the pressure wave into an
electrical signal. A well-functioning flush system is essential for obtaining undistorted waveforms
and accurate information.
IP monitoring is available on either Channel 2 or 3. The IP monitor is compatible with industry
standard (IEC 60601-2-34) pressure transducers with 5µV/V/mmHg sensitivity. The transducer
must provide defibrillation protection of at least 360 joules. The following IP transducers may be
used with the LIFEPAK 15 monitor/defibrillator.
MANUFACTURER
DESCRIPTION
Utah Medical
DPT-100 Deltran
®
Disposable Pressure Transducer
Edwards Lifesciences
TruWave™ Disposable Pressure Transducer
ICU Medical
Transpac
®
IV Disposable Pressure Transducer
An invasive pressure adapter cable is used to connect the transducer to the monitor. The IP
connector is a 6-pin type 3102A-14S-6S connector. The connector pinout has the following
configuration, counterclockwise from 12 o’clock, viewed from the front of the
LIFEPAK 15 monitor/defibrillator.
A pin = - signal
B pin = + excitation
C pin = + signal
D pin = - excitation
E pin = shield
F pin = unlabeled
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......