Chapter 7
| Data Management
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
165
EVENT NAME
WAVEFORM DATA (WHEN CAPTURED)
PACING X PAUSED
Initial 8 seconds while
PAUSE
is pressed
ALARM*
3 seconds prior to violated parameter and 5 seconds after
EVENT*
3 seconds prior to event selection and 5 seconds after
3 seconds prior to pressing
and 5 seconds after
12-LEAD
10 seconds after
12-LEAD
is pressed
SNAPSHOT
3 seconds prior to and 5 seconds after SNAPSHOT requested
VITAL SIGNS
3 seconds prior to and 5 seconds after vital signs are acquired
*To reduce the length of the CODE SUMMARY report, storing waveform data with these events can be set to OFF (see
the
LIFEPAK 15 Monitor/Defibrillator Setup Options
provided with your device).
Waveform events are preceded by a header that includes the following information:
•
Patient data
•
Vital signs
•
Event name
•
Device configuration information
•
Therapy data*
*Patient impedance (in ohms) appears on shock reports when using disposable defibrillation electrodes. This
impedance is measured just prior to the shock and is used to determine voltage compensation.
The following figures show four examples of waveform events as they would appear in the
CODE SUMMARY report.
Analysis Event
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......