General Troubleshooting Tips
214
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
General Troubleshooting Tips
If a problem is detected with the LIFEPAK 15 monitor/defibrillator during operation or testing,
refer to the following troubleshooting tips. If the problem cannot be corrected, remove the
LIFEPAK 15 monitor/defibrillator from active use and contact a qualified service technician for
service and repair.
Table 37
General Troubleshooting Tips
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
No power when
monitor/defibrillator is
turned
ON
Low battery voltage
•
Replace with fully charged, properly
maintained battery.
Battery connector pin loose,
covered with foreign
substance, or damaged
•
Remove battery and inspect pins.
Clean if foreign substance present.
Contact a qualified service
technician to replace if bent,
cracked, or loose.
Power adapter not properly
connected to auxiliary power
source
•
Check that power adapter is
properly connected to auxiliary
power.
Power adapter not properly
connected to
monitor/defibrillator
•
Check that power adapter is
properly connected to
monitor/defibrillator.
Defective power adapter or
cables
•
Replace with working power adapter
and cables.
Defective battery
•
Remove battery from service and
replace with working battery.
ON
LED illuminated, but
screen is blank and device
does not operate
Device boot up has failed
•
Press and hold
ON
until LED turns
off (~5 seconds). Then press
ON
to
turn device back on.
•
If device does not turn off, remove
both batteries and disconnect
device from power adapter, if
applicable. Then reinsert batteries,
reconnect power adapter, and press
ON
to turn device back on.
AUXILIARY POWER
LED
not illuminated
Power adapter not properly
connected to auxiliary power
source
•
Check that power adapter is
properly connected to auxiliary
power.
Power adapter not properly
connected to
monitor/defibrillator
•
Check that power adapter is
properly connected to
monitor/defibrillator.
Defective power adapter or
cables
•
Replace with working power adapter
and cables.
BATTERY CHARGING
LED
on monitor/defibrillator not
illuminated
Power adapter not properly
connected to auxiliary power
source or
monitor/defibrillator
•
Check that power adapter is
connected properly.
Battery not properly inserted
in battery well
•
Check that battery is properly
inserted in battery well.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......