Specifications and Performance Characteristics
234
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Paddle Options
QUIK-COMBO pacing/defibrillation/ECG electrodes (standard)
Standard paddles (optional)
Cable Length
8 foot long (2.4 m) QUIK-COMBO cable (not including electrode
assembly)
AED Mode
Shock Advisory System (SAS) is an ECG analysis system that advises
the operator if the algorithm detects a shockable or nonshockable
ECG rhythm. SAS acquires ECG via therapy electrodes only.
Biphasic Output Energy Shock levels ranging from 150–360 joules with same or greater energy
level for each successive shock
cprMAX™ Technology
In AED mode, cprMAX technology provides a method of maximizing
the CPR time that a patient receives, with the overall goal of improving
the rate of survival of patients treated with AEDs
Setup Options:
Auto Analyze Allows for auto analysis. Options are
OFF
,
AFTER 1ST SHOCK
Initial CPR Allows the user to be prompted for CPR for a period of time prior to
other activity. Options are
OFF
,
ANALYZE FIRST
,
CPR FIRST
Initial CPR Time Time interval for Initial CPR. Options are
15
,
30
,
45
,
60
,
90
,
120
,
and
180
seconds
Pre-Shock CPR Allows the user to be prompted for CPR while the device is charging.
Options are
OFF
,
15
,
30
seconds
Pulse Check Allows the user to be prompted for a pulse check at various times.
Options are
ALWAYS
,
AFTER SECOND NSA
,
AFTER EVERY NSA
,
NEVER
Stacked Shocks Allows for CPR after 3 consecutive shocks or after a single shock.
Options are
OFF
,
ON
CPR Time 1 or 2 User selectable times for CPR. Options are
15
,
30
,
45
,
60
,
90
,
120,
180
seconds and
30
minutes
PACER
Pacing Mode
Demand or nondemand
Rate and current defaults
Pacing Rate
40 to 170 PPM
Rate Accuracy
±1.5% over entire range
Output Waveform
Monophasic, truncated exponential current pulse (20 ±1 msec)
Output Current
0 to 200 mA
Pause: Pacing pulse frequency reduced by a factor of 4 when
activated
Output Current
Accuracy
± 10% or 5 mA (whichever is greater) over the specified load
impedance range
Refractory Period
180 to 280 msec (function of rate)
Physio-Control Therapy
Electrode Post-Pacing
Performance per
IEC 60601-2-4
After pacing:
AC large signal impedance
≤
4.2
Ω
DC offset voltage
≤
1053 mV
After pacing followed by 360 J shock:
DC offset voltage
≤
1228 mV, 4 seconds after shock
DC offset voltage
≤
966 mV, 60 seconds after shock
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......