Appendix A
| Specifications and Performance Characteristics
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
227
Heart Rate with Irregular
Rhythm
The rate meter output can range from the heart rate associated with
the shortest R-R interval to the heart rate associated with the longest
R-R interval. When present, intermediate length R-R intervals are
favored as the basis for the rate.
When evaluated per IEC 60601-2-27, rates are as follows:
•
A1. Ventricular bigeminy: HR = 80 to 86
•
A2. Slow alternating ventricular bigeminy: HR = 60 to 63
•
A3. Rapid alternating ventricular bigeminy: HR = 123 to 124
•
A4. Bidirectional systoles: HR = 97 to 99
QRS Detection Range
Duration: 40 to 120 msec
Amplitude: 0.5 to 5.0 mV
Tall T-wave Rejection: T-waves that are 1 mV high are not detected by
the monitor when the R-wave size is 1 mV and input rate is 80 ppm.
Common Mode
Rejection (CMRR)
ECG Leads: 90 dB at 50/60 Hz
Pacemaker Pulse
Rejection
Rejects pacemaker pulses having amplitudes from ±2 mV to ±700 mV
and pulse widths from 0.1 ms to 2.0 ms with and without overshoot.
Pacemaker pulse overshoot is defined as 2.5% to 25% of the
pacemaker pulse amplitude not to exceed 2 mV. Refer to IEC 60601-
2-27.
Note:
Does not apply when internal paddles are connected.
SpO
2
/SpCO/SpMet
Sensors
Masimo
®
sensors including Rainbow
®
sensors
Nellcor
®
sensors when used with the Masimo Red™ MNC adapter
SpO
2
Displayed Saturation
Range
"<50" for levels below 50%; 50 to 100%
Saturation Accuracy is specified for range 70-100% (0-69% is not specified).
Adults/Pediatrics
Accuracy (RMS)*
±2% (during no motion conditions - Masimo)
±3% (during no motion conditions - Nellcor)
±3% (during motion conditions - Masimo)
Neonatal Accuracy
(RMS)*
±3% (during no motion conditions - Masimo)
±4% (during no motion conditions - Nellcor
±3% (during motion conditions - Masimo)
Dynamic signal strength bar graph
Pulse tone as SpO
2
pulsations are detected
SpO
2
Averaging Time
User selectable: 4, 8, 12 or 16 seconds
SpO
2
Data Update
Period
1 second
SpO
2
Alarm Condition
Delay
21 seconds (maximum time from physiological change to detection
by device, with default 8 second averaging selected)
SpO
2
Alarm Signal
Generation Delay
1 second (time for device to generate alarm after alarm condition is
detected)
SpO
2
Sensitivity
User selectable: Normal, High
SpO
2
Measurement
Functional SpO
2
values are displayed and stored
Pulse Rate Accuracy (specified for range 25 to 240 bpm)
Adults/Pediatrics
(RMS)
±3% (during no motion conditions)
±5% (during motion conditions)
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......