Specifications and Performance Characteristics
224
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Analog ECG Output
Output: 1 volt/mV
Frequency Response: 0.67 to 32 Hz (except 2.5 to 25 Hz for Paddles
ECG and 1.3 to 23 Hz for 1–30 Hz Monitor Frequency Response)
No internal pacemaker pulse enhancements or detection markers
included. Pace markers during external pacing using this device are
included.
Notch Filter
50 or 60 Hz
POWER
Batteries
Rechargeable Lithium-ion battery
Dual battery capability with automatic switching
Low battery indication and message: Low battery fuel gauge
indication and low battery message in status area for each battery
Replace battery indication and message: Replace battery fuel gauge
indication, audio tones, and replace battery message in the status
area for each battery. When replace battery is indicated, device auto-
switches to second battery. When both batteries reach replace battery
condition, a voice prompt instructs user to replace battery.
Input voltage range is b8.8 and +12.6 Vdc
Battery Capacity
For two, new fully-charged batteries, 20°C (68°F):
Capacity to shutdown is:
Operating Mode
Monitoring (minutes)
Pacing (minutes)
Defibrillation
(360J discharges)
Typical
360
340
420
Minimum
340
320
400
Capacity after low battery is:
Typical
21
20
30
Minimum
12
10
6
AC Power Adapter
AC-DC power adapter
Input power range is 100-240 Vac, 50/60 Hz, 1.4-0.6 A
Output voltage is 12 Vdc
Meets UL 60601-1 300 microampere Earth leakage limit when
installed on a center-tapped, 240 Vac, single phase circuit.
DC Power Adapter
DC-DC power adapter
Input power range is:
Minimum: 11 Vdc, 15 A
Nominal: 13.8 Vdc, 12.5 A
Maximum: 17.6 Vdc, 10 A
Output voltage is 12 Vdc
Device Behavior
when using
Power Adapter
Auxiliary power indicator on defibrillator illuminated when connected
to auxiliary power. Battery charging indicator illuminated when
batteries are fully charged and flashing if either battery is being
charged.
Battery status indicators on display show battery charge level, but well
number is not highlighted because battery is not in use. Low battery
and replace battery prompts and messages do not occur.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......