Appendix A
| Specifications and Performance Characteristics
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
233
Waveform Shape and Measured Parameters
Biphasic Waveform
Patient
Impedance (
Ω
)
Phase 1 Duration (ms)
Phase 2 Duration (ms)
Tilt (%)
Min
Max
Min
Max
Min
Max
25
5.1
6.0
3.2
4.2
69.9
85.2
50
6.8
7.9
4.4
5.5
57.0
74.7
75
7.6
9.4
4.9
6.5
49.3
67.6
100
8.7
10.6
5.6
7.3
43.0
62.2
125
9.5
11.2
6.2
7.7
39.0
56.6
150
10.1
11.9
6.6
8.2
36.8
52.6
175
10.6
12.5
6.9
8.6
33.8
49.3
Rated Energy Output
Rated energy output is the nominal delivered energy based on the energy setting and patient
impedance, as defined in the following chart.
*Energy setting selected
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......