Monitoring SpO2, SpCO, and SpMet
78
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
High intensity lights (such as
pulsating strobe lights) may be
interfering with performance
•
Cover sensor with opaque
material, if necessary.
Damaged cable or sensor
•
Replace damaged cable or
sensor.
Different SpCO or SpMet
measurements on same patient
Every measurement, even on
the same patient, can be
different
•
Confirm by taking three
measurements: ring finger,
middle finger, and then index
finger; average the results.
XXX
appears in place of
SpO
2
reading
SpO
2
module failed.
Internal cable failed.
•
Turn device off and then on
again.
•
If problem persists, contact
qualified service personnel.
SPO2: CHECK SENSOR
message appears
Sensor is disconnected from
patient or cable
•
Attach the sensor.
•
Check that sensor is secure.
Excessive ambient light
•
Remove or block light
source, if possible.
•
Cover sensor with opaque
material, if necessary.
Faulty or defective sensor
•
Replace sensor.
Patient has a weak pulse or low
blood pressure, or the sensor is
not properly placed
•
Change sensor location.
•
Check if patient perfusion is
adequate for sensor location.
•
Check that sensor is secure
and not too tight.
•
Check that sensor is not on
extremity with blood
pressure cuff or intravascular
line.
•
Test sensor on someone
else.
SPO2: UNKNOWN SENSOR
message appears
A sensor that is not
Physio-Control approved is
connected to the device
•
Check that the sensor is
approved by Physio-Control.
•
If using Nellcor sensor,
check that it is connected to
monitor using Masimo
Red MNC adapter cable.
SPO2: SEARCHING FOR PULSE
message appears
A sensor is connected to the
patient and is searching for a
pulse
•
Wait for completion.
SPO2: LOW PERFUSION
message appears
Patient has a weak pulse
•
Change sensor location.
SP02: POOR QUALITY SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
•
Check that sensor and cable
are connected properly.
•
Move sensor to a better
perfused site.
SPCO: POOR QUALITY SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
•
Check that sensor and cable
are connected properly.
•
Move sensor to a better
perfused site.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......