Chapter 10
| Maintaining the Equipment
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
215
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Unable to charge battery
with power adapter because
battery charge level is too
low.
•
Charge battery in Station-Mobile or
REDI-CHARGE battery charger if
available.
•
Replace battery.
No batteries installed
•
Install at least one battery.
Defective battery
•
Remove battery from service and
replace with working battery.
Unrecognized battery
•
Only use battery that is approved for
use with the
LIFEPAK 15 monitor/defibrillator.
Incompatible power adapter
connected to the
monitor/defibrillator
•
Only use power adapter that is
approved for use with the
LIFEPAK 15 monitor/defibrillator.
Defective power adapter or
cables
•
Replace with working power adapter
and cables.
Monitor/defibrillator unable
to recognize installed battery
•
Contact qualified service personnel.
CANNOT CHARGE
BATTERY
message
appears
Defective battery
•
Remove battery from service and
replace with working battery.
Defective power adapter
•
Replace with working power adapter
and cables.
Device unable to charge
battery or batteries
•
Contact qualified service personnel.
Fuel gauge on battery does
not illuminate
Extremely depleted battery
•
Charge battery in Station-Mobile or
REDI-CHARGE battery charger.
Faulty battery
•
Replace battery.
Device turns off
unexpectedly
High power draw
•
Press
ON
immediately to turn device
back on.
Low battery power
•
Replace battery immediately.
•
Press
ON
to turn device back on.
RF equipment too close to
defibrillator
•
Separate RF equipment from
defibrillator. See Separation
Distances (on page 266).
•
Press
ON
to turn device back on.
Cellular equipment too close
to installed battery
•
Move cellular equipment away from
installed battery.
•
Press
ON
to turn device back on.
•
If device does not turn on, replace
battery.
LIFENET Gateway (modem)
too close to installed battery
•
Store modem in side pouch of
defibrillator. Do not store modem in
back pouch.
•
Press
ON
to turn device back on.
•
If device does not turn on, replace
battery.
Device won’t turn off
ON
not pressed long enough
to turn off device
•
Press and hold
ON
for at least two
seconds.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......