Appendix B
| Screen Messages
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
245
Summary of Screen Messages
Table 44
Summary of Screen Messages
MESSAGE
DESCRIPTION
12-LEAD ECG UNAVAILABLE
A 12-lead was requested but the necessary ECG data is not
available.
ABNORMAL ENERGY DELIVERY
A discharge occurred when the paddles were shorted
together, when hard paddles did not have adequate contact
with the patient or were discharged in the air, or patient
impedance was out of range. Message may also appear in
certain types of internal faults.
ACCESS DENIED
Three consecutive incorrect passcode attempts were made to
enter Manual mode.
ACQUIRING 12-LEAD
Monitor is acquiring data for 12-lead ECG report.
ACQUIRING SNAPSHOT
A snapshot report of current vital signs has been requested.
ADVISORY MODE-MONITORING
The device is monitoring the patient ECG for a shockable
rhythm.
ADVISORY: SPCO > 10%
SpCO advisory alert activated. SpCO value is greater
than 10%.
ADVISORY: SPMET > 3%
SpMet advisory alert activated. SpMet value is greater
than 3%.
ALARM APNEA
No valid breath has been detected for 30 seconds.
ALARMS SILENCED
Alarms are silenced. An alert tone with status message
ALARMS SILENCED
occurs periodically as a reminder.
ANALYZING 12-LEAD
The data for 12-lead ECG report is being analyzed.
ANALYZING NOW-STAND CLEAR The AED is analyzing the patient ECG rhythm.
ATTEMPTING TO TRANSMIT
The device is processing a transmission request.
AUTO NIBP CANCELLED
The automatic initiation of NIBP measurements has been
cancelled.
BATTERY X LOW
The specified battery has a low energy condition.
BLUETOOTH DEVICE NOT FOUND
Bluetooth
device has not been detected.
BLUETOOTH UNAVAILABLE
Unable to locate or connect to target device.
CANNOT CHARGE
CHARGE
is pressed and the synchronize source is missing for
synchronized cardioversion, the therapy cable is not
connected, or QUIK-COMBO electrodes are not attached to
the therapy cable.
CANNOT CHARGE BATTERIES
Both batteries are installed, and the device is unable to charge
either battery.
CANNOT CHARGE BATTERY 1
The device is unable to charge the battery in battery well 1.
CANNOT CHARGE BATTERY 2
The device is unable to charge the battery in battery well 2.
CHARGING TO XXX J
Appears when
CHARGE
is pressed on the front panel or
standard paddles.
CHECK FOR PULSE
AED prompt after each standard 3-shock sequence or
NO
SHOCK ADVISED
message.
CHECK PATIENT!
A potentially shockable rhythm has been detected when the
VF/VT alarm is on.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......