GB 04373 - Edition 01 - Feb. 10
1/1
B 85
Alcatel Vacuum Technology France - ATH 200 User’s Manual
RS 232 / RS 485 / Network interface
installation
Disconnect the pump from the electrical connection
E 14
§ turning off the pump.
Study the preliminary precautions
B 00
.
•
Preliminary operations
3
Loosen and remove the 4 M4x8 Chc screws holding the network
blank off plate on the pump rear panel.
Cut the collar holding the interface supply cable on the network blank
off plate.
Connect the pump interface supply cable plug on the RS 232 / RS 485
/ Network interface connector.
Insert this connector interface instead of the network blank off plate.
3
Tighten the 4 M4x8 Chc screws holding the RS 232 / RS 485 /
Network interface on the pump rear panel.
Connect the pump power supply
B 40
.
Note : When the main switch is in position 1, the led in front face of the RS 232 /
RS 485 / Network interface lights up.
RS 232 / RS 485 /
Network interface
Network blank off plate
Rear panel
Chc screw
RS 232 / RS 485 / Network interface connector
Chc screw
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions