GB 04358 - Edition 03 - Dec. 11
1/4
B 40
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Electrical connection
Make sure that main switch is on position «0» during electrical
connection.
Study the preliminary precautions
B 00
.
All the internal electrical connections required for use the pump are
made prior to shipment. However, the electrical connection of the main
power supply is provided by the process tool.
A100L includes 2 medium time lag to protect the electronic cabinet:
they aren’t operator replaceable fuses.
•
•
Protective earth
connection
Pump
Voltage
Medium time lag
A100L
380/460 V 50/60 Hz
0.4A/ 500 Vac
A100L
A100L ES
200/208 V 50/60 Hz
0.1A/250 Vac
Electric shock hazard.
The voltages and currents in use can induce electric shock.
Isolate and lock out power line by switching off the main isolator before
maintaining the product /or removing the cover.
This main isolator has an interrupting short circuit current of 10kA.
Take care! items located between the mains connection and the isolator are
still under mains voltage: disconnect the mains cable from all power sources
before commencing any maintenance work on the product.
Only skilled, authorized people may carry out maintenance work.
Ensure that the equipment and the electrical supply cable are suitably
protected against earth defects and that the earth wire is longer than the
three conducting wires.
The user must provide a second protective earth (ground) conductor
with a cross-sectional area at least equal to the conducting wire size. Than,
connect the protective earth stud to the suitable installation earth point.
An IEC 417#5017 symbol
is located inside the enclosure for each
ground terminal.
An IEC 417#5019 symbol
is located on the front panel of the
pump under the main power supply for the main power supply ground
terminal
A 40
.
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions