GB 04365- Edition 02 - Dec 11
1/1
D 10
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Pump operating status information
Study the safety instructions related to maintenance
D 00
.
•
Operation monitoring
During operation, the user is warned of an operating incident by:
Indicator lights on the front panel.
Green on
: pump is running and no problems
Yellow and green on
: warning
Red on
: hazard
Pumping is stopped when an hazard threshold is reached.
The fault contacts on the remote control connector at the rear of the
pump are open.
Maintenance
operations
section E
Hazard
Red light
Warning
Yellow light
ON
Green light
Defect status
Warning status
The yellow light is lighting up.
The pump is not stopped.
Hazard status
The red light is lighting up.
The pump is stopped.
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions