GB 04732 - Edition 01 - Dec 11
1/1
B 35
Alcatel Vacuum Technology France - ATH 200 User’s Manual
ES (Energy Saving) option
Inert gas N
2
or CDA connection
Nitrogen or CDA
characteristics
For maximum performances,
a filtered dry nitrogen supply
with the
following characteristics is required:
N
2
: 78.09 %
O
2
: 20.93 %
H
2
: 0.01 %
Flowrate
A 40
Pressure range
0 to 7 bar absolute (0 to 101 psi)
Min. pressure
2 bar / 7 bar absolute (15 / 87 psig)
Connection
Connect the gas line on the
N
2
/CDA connector
.
Rear panel
Swagelok for tube 6.35 mm
Smooth pipe
external diameter 6,35 mm
Customer supply
Manufacturer supply
A sudden inrush of N
2
or CDA into the pressurization circuit may result in
damage to internal parts.
We advise to install a pressure regulator at the gas line.
Close the pressure regulator before pressurizing the circuit.
Inlet CDA or N
2
connection option