2/5
C 55
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Use of the RS 232 serial link
GB 04375 - Edition 02 Feb. 11
List of controls
Order
Paramater
Response
Action
ADR
aaa
OK
or
ERR2
Specifies the address of the device for networking.
adr =
address of the device before the command
aaa =
new address of the device
condition : 000 aaa 255
•
ECH
ON or OFF
OK
or
ERR2
Enables or disables command echoing.
•
IDN
None
V X.ZZ
Identifies the device which is communicating, and its software
version.
Example : #adr,M20A100 V1.04
0 :
Header
1 to 3 :
Address
4 :
Separator
character
5 to 19 :
Type of monitoring
20 to 24 :
Software version (x), software edition (zz)
25 :
Carriage
return
•
LEV
None
Returns the state of the parameters defined by SET.
Example: #adr,A001,A002,B001,B002,C001,C002,D001,D002,E001,E002,F001,F002,G000,
H0000,I0000
,K0000,L0000,M0000,N0000,O001,O002,P001,P002,Q001,Q002<CR>
0 :
: Header
1 to 3
: Address
4 :
Separator
character
5 to 68
: Reserved
69 :
Separator
character
70 to 74
: Full overhaul (x100h)
75 :
Separator
character
76 to 140 : Reserved
141 :
Carriage
return
•
SEL
None
Returns the state of the parameters defined by OPT.
Example: #adr,A,B,C,D,E,F,G,H,I,J,K,L,M,N,O,P,Q,R,S,T<CR>
0
: Header
1 to 3
: Address
4 :
Separator
character
5 to 11
: Reserved
12 :
Separator
character
13
:
System selection
: 5 = A100L/A180L
14 :
Separator
character
15
:
Power start
: 0 = Valid / 1 = Disabled
16 to 27 :
Reserved
•
List of controls
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions