Alcatel Vacuum Technology France - ATH 200 User’s Manual
Electrical connection
GB 04358 - Edition 03 - Dec. 11
Process tool electrical
protection
Pump
Voltage
Circuit breaker rating
A100L
380/460 V - 50/60 Hz
15A
A100L
A100L ES
200/208 V - 50/60 Hz
15A
The user must supply the pump from facilities equipped with main cir-
cuit breaker, curve D (IEC 60947-2), in accordance with local regula-
tions and with a minimum interrupting short circuit current of 10 kA.
This protection device must be in close proximity to the pump
(no further than 7m (25 ft)) easily accessible by an operator, within
line of sight of the pump. Besides, mark it visibly as an electrical
interrupting device of the product.
For any change of power supply voltage in relation to the initial
configuration, contact our manufacturer’service center.
Customer electrical
installation protection
The circuit breaker mustn’t switch off the ground conductor, nevertheless it
must protect against overcurrent.
Installation protection with
circuit breaker