Maintenance
83
WARNING
AVERTISSEMENT
Risk of burns. Never touch a heated mass spectrometer transfer
line or a GC injector cap with unprotected (bare) fingers.
Risque de brûlures. Ne touchez jamais une ligne de transfert de
spectromètre de masse chauffée ou un capuchon d'injecteur GC
avec des doigts non protégé.
4.
Pull the inner transfer line tube back 25 mm (1 inch).
1/4-inch Nut
Column
Inner Transfer Tube
Pull Back
Figure 9
Pulling back the inner transfer tube.
5.
Open the Clarus SQ 8 MS access door.
Содержание Clarus SQ 8 MS Series