Clarus SQ 8 MS Hardware Guide
116
2.
Remove the source. See
Removing and Returning the Source
on page
3.
Use the medium sized Philips screwdriver to remove the two screws on the left
access panel. See the following figure.
Figure 30 Removing the cover screws.
4.
Remove the left access panel and place in a secure location.
Screws