Maintenance
121
13.
From the front of the optics assembly disconnect the fan cable connector,
reference gas valve cable connector, the source connector and the nut for the
reference gas vacuum hose. See the following figure.
Figure 36 Disconnecting the cables at the front of the optics assembly.
14.
Hold the ion optics assembly by its handles, lift it up, and place it on a clean
surface.
15.
Cover the vacuum manifold with aluminum foil.
Reference Gas Valve
Cable Connector
Fan Cable
Source
Connector
Nut and Ferrule for
Reference Gas Vacuum
Hose