Clarus SQ 8 MS Hardware Guide
152
Problem
Probable Cause
Solution
The ultimate pressure is
poor.
Is the cooling inadequate?
Check the cooling-air flow and
correct if possible.
Check the cooling-air duct for
obstructions and correct as
necessary.
If the cooling air flow is fine
and there are no obstructions
contact your PerkinElmer
service representatives.
Is the backing pressure
high?
Check for a leak in the backing
pipeline and poor backing
pump performance. Correct as
necessary.
The pump is very noisy or
there is excessive vibration
or both.
Is the noise irregular and
getting progressively
worse?
If so, a bearing may be
defective. Contact your
PerkinElmer service
representative.
Is the pump making a
constant high pitched
noise?
If so, the rotor may be out of
balance. Contact your
PerkinElmer service
representative.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......