Troubleshooting
153
Problem
Probable Cause
Solution
Turbo pump will not
accelerate
Pump malfunction.
Call a PerkinElmer service
engineer.
Pump controller
malfunction.
Call a PerkinElmer service
engineer.
GC is not properly
configured.
Set the proper GC
configuration for your site.
Vacuum light continues to
blink.
Large leak.
Locate vacuum leak and
correct.
Foreline trap has
excessive moisture.
Replace filter.
Rotary pump set to gas
ballast.
Switch the gas ballasting off.
Rotary pump requires oil
change.
Change oil.
High mass spectra appears
as a large blotch, or loss of
high mass spectra.
Bad tune.
Run AutoTune.
Drastic change in mass peak
shape for no apparent
reason.
Bad tune.
Run AutoTune.
No spectra, or large blotch.
Bad tune.
Run AutoTune.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......