Clarus SQ 8 MS Hardware Guide
144
keep the source cleaner, but may increase
fragmentation.
For example, down a little from EI to minimize
fragmentation. 120
°
C is the practical lower limit.
Ion Energy
Approx. 1 or 2
2.
Optimize the amount of reagent gas flowing into the source by using two
heptacosa ions, m/z 452 and 633, which usually produce relative intensities of
65 – 85% and 95% respectively.
Heptacosa can be used to calibrate the m/z range for negative ion CI analyses.
3.
Maximize the peak intensities, then slightly decrease the reagent gas by turning
the knob clockwise 1/8 turn.
Optimize the tuning parameters for maximum intensity.
4.
Save the Tune page parameters by selecting
Save As…
from the
File
menu.
5.
Select
Calibrate Instrument
from the Tune page
Calibration
menu.
6.
Select
heptaneg.ref
from the drop-down menu.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......