Clarus SQ 8 MS Hardware Guide
162
Chromatography Related
Problem
Probable Cause
Solution
Inconsistent retention
time.
Injector septum leak.
Replace the septum.
Carrier gas manifold leak.
Locate and fix the leak.
Rising Total Ion
Chromatogram baseline.
Column bleed.
Disconnect the column from
the mass spectrometer and
condition the column.
Vacuum leak.
Locate and fix the leak.
Discreet high intensity
contaminant peaks.
Column bleed.
Disconnect the column from
the mass spectrometer and
condition the column.
Injector septum bleed.
Replace the septum and/or
glass liner.
Tailing peaks (sloping on
RHS).
Improperly installed column.
Check the column and
reinstall if necessary.
Injector too cool.
Raise the injector
temperature.
Interface temperature too cool. Raise the interface
temperature.
Inadequate carrier gas flow.
Set proper flow.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......