Maintenance
143
8.
Tuning may be optimized on the m/z 69, 219, 414, and 652 ions of the
heptacosa reference gas.
9.
Click
Press for Standby
to turn off Operate, followed by the CI gas.
You are now ready to run your CI+ analysis.
Adjust the Reagent Gas for CI-
1.
Open the CI gas inlet by selecting CI Gas from the Gas menu. Wait at least 10
seconds before clicking
Press for Operate
.
Parameter:
CI
−
value and comments:
Electron Energy
30 to 70 eV
(This parameter should be optimized.)
Emission
200 to 300
µ
A is acceptable
(Emission measures the real emission current, i.e.
the source current from the source block, there is no
separate measurement of source current in CI.)
Lens 1 and 2
The tuning of these lenses may be
different from the optimum values set for
EI, since the source pressure is much
higher in CI.
Multiplier
1300V to 1600V
Source temperature
150
°
C is standard. Higher temperatures
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......