Clarus SQ 8 MS Hardware Guide
166
Spectral Related
Problem
Probable Cause
Solution
Noisy spectra.
Dirty source.
Remove and clean the source.
Peak detection threshold set too low.
Raise the thresholds.
Multiplier voltage set too high.
Lower the multiplier voltage.
Spectrum distortion.
Scanning too fast or slow.
Reset the scan rate.
Incorrect Isotope ratios. Poor calibration.
Recalibrate.
Incorrect tune.
Retune.
Defective filament.
Replace filament.
Air leak.
Find and fix.
Missing Isotopes in
spectrum.
Bad calibration.
Recalibrate.
Poor tuning.
Retune.
Dirty source.
Clean source.
Sample too weak.
Use a higher sample
concentration.
Peak detection thresholds set too
high.
Lower the thresholds.
Multiplier voltage set too low.
Raise the multiplier voltage.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......