Bench Space
The laboratory bench should be sturdy enough to support
the full weight of the GC/MS as well as additional
equipment (for example, computer and/or printer). Expect
the total weight of the GC/MS and accessory equipment to
weigh at least 159 kg (350 lb).
Allow a minimum clearance of 15 cm (6 in.) on each side,
22.9 cm (9 in.) at the rear, and 137.2 cm (54 in.) at the top
of the GC/MS. If this is not possible, install the GC/MS on
a bench that has wheels. The bench requires an area
underneath for the forepump.
Do not position the Clarus SQ 8 T and C so that it is
difficult to operate the AC power on/off switch on the
lower left side of the instrument in case of a malfunction of
the instrument. For the Clarus SQ 8 S the AC power on/off
switch is on the back of the instrument.
Environmental Requirements
Pollution Degree
This instrument will operate safely in environments that
contain non-conductive foreign matter up to Pollution
Degree 2 as defined in EN/IEC 61010-1.
Laboratory Environment
Install the GC/MS in an indoor laboratory environment that
is clean and free of drafts and direct sunlight.
The laboratory should be free of flammable, explosive,
toxic, caustic or corrosive vapors or gases, and should be
relatively free of dust.
The ambient laboratory temperature should be between 10
°C and 30 °C (50 °F and 86 °F) for Clarus SQ 8 C, SQ 8 T,
and SQ 8 S systems unless the turbomolecular pump is
water cooled, and between 10 °C and 35 °C (50 °F and 95
°F) for Clarus SQ 8 C, SQ 8 T, and SQ 8 S systems with
water cooling.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......