Troubleshooting
157
Problem
Probable Cause
Solution
Poor sensitivity (correct
amount of sample is reaching
the mass spectrometer).
Column is improperly
positioned in the ion
source.
Reinstall the column and
check the cut at the end
that fits in the source.
Improper tuning, or a
dirty or defective ion
source.
Check the tuning. Increase
the multiplier voltage.
Shut down the system,
remove the inner source,
clean or replace lenses in
the outer source, and
install a new filament.
Loss of resolution (especially
at high mass).
Dirty prequadrupole rods.
Clean the prequadrupole
rods.
Mass assignment drifts.
Large temperature
fluctuations in the
laboratory.
Stabilize the lab
temperature, or isolate the
GC/MS system from large
temperature fluctuations.
Mass assignment incorrect.
Run mass calibration.
Skewed spectra.
Improper scan speed (too
slow for the narrow peaks
produced by capillary
columns).
Increase the scan speed.
Unusually high repeller
voltage.
Repeller dirty.
Clean repeller.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......