Clarus SQ 8 MS Hardware Guide
130
3.
Run the pump until the oil appears clear.
Changing the Oil
1.
Warm the oil by running the pump for at least 10 minutes, and then switch off
the vacuum system.
2.
Unplug the pump from the AC outlet and disconnect it from your vacuum
system.
3.
Remove one of the oil filler plugs.
Figure 43 Draining forepump oil.
Filler Plug
Drain Plug
2 Liter
Drain Container
Max Oil Level