Troubleshooting
163
Problem
Probable Cause
Solution
Dirty injector liner.
Clean or replace.
Column has active sites.
Equilibrate or replace.
Chromatographic peaks
too wide.
Injector too cool.
Raise injector temperature.
Sample overloading the
column.
Use a split injection or a smaller
sample.
Incorrect GC oven program.
Enter a new oven program.
Discrimination of relative
peak intensities.
Poor resolution or improper
tuning.
Retune.
Unstable filament.
Replace filament.
Poor calibration.
Recalibrate.
Air leaks at detector.
Check He/Air ratio.
Peaks are flat-topped.
Signal strength exceeds
dynamic range of detector.
Reduce multiplier voltage.
Sample is too strong.
Dilute or split.
High baseline.
Dirty sample.
Prepare and filter a new sample.
Air leak at injector.
Locate and fix the air leak.
Contaminated carrier gas.
Replace the gas tank.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......