Maintenance
77
To display the Tune page:
1.
While displaying TurboMass sample list screen click
.
The TunePage displays.
2.
Turn on the filament and high voltages by clicking
Press for Operate
at the
bottom right of the window.
The indicator box turns green to indicate that it is on.
3.
Select
UltraTune/Custom (AutoTune)
from the
Options
menu, then click
Start
.
You will hear a click when the reference gas solenoid valve opens and
AutoTune begins. Upon completion, the message
AutoTune completed
successfully
is displayed.
4.
Click
OK
.
5.
Select
Reference Gas On
from the
Gas
menu to remove the check mark (
),
or click
to set it in the
up
position.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......