Clarus SQ 8 MS Hardware Guide
160
Problem
Probable Cause
Solution
Piece of column broken off in
the ion chamber.
Remove the inner source,
check for and remove piece of
column.
Source filament is bent.
Check and replace filament if
necessary.
RF generator malfunction.
Call a PerkinElmer service
engineer.
Analyzer drive electronics
malfunction.
Call a PerkinElmer service
engineer.
Total Ion Chromatogram
too high.
Dirty source.
Clean the source.
Contamination from poor
handling technique.
Set source and transfer line to
250 ºC and maintain this
temperature overnight.
Stationary phase of column
de-polymerizing (bleeding).
Change column.
Air leak.
Find the leak and fix it.
Poor quality carrier gas.
Replace the carrier gas tank.
Carrier gas filter is ineffective
and needs replacing.
Replace the carrier gas filter.
No noise on mass
chromatogram.
Detector multiplier voltage too
low.
Increase the multiplier
voltage.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......