Clarus SQ 8 MS Hardware Guide
168
Communications Related
Problem
Probable Cause
Solution
Will not boot MS.
PC (computer) to MS cable
has a loose connection.
Check and reset the cable.
Transient in power supply
has halted communications.
Reboot the PC (computer).
Will not control GC.
RS 232 communications
cable loose connections.
Check and restart the mass
spectrometer.
Power failure/transient
surge to GC or
autosampler.
GC electronic malfunction.
Call a PerkinElmer service
engineer.
Communication cable
intermittent contact.
GC electronics
malfunction.
Call a PerkinElmer service
engineer.
Crashes when starting an
acquisition.
Software corrupted.
Reload software.
Rotary pump malfunction.
Call a PerkinElmer service
engineer.
Содержание Clarus SQ 8 MS Series
Страница 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus SQ 8 MS Hardware Guide ...
Страница 8: ......
Страница 9: ...Warnings and Safety Information 1 ...
Страница 10: ......
Страница 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Страница 46: ......
Страница 47: ...Introduction 2 ...
Страница 48: ......
Страница 52: ......
Страница 53: ...About the Clarus GC MS System 3 ...
Страница 54: ......
Страница 64: ......
Страница 67: ...Maintenance 65 Maintenance 4 ...
Страница 68: ......
Страница 69: ...Maintenance 67 ...
Страница 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Страница 148: ...Troubleshooting 5 ...
Страница 149: ...Troubleshooting 147 ...
Страница 178: ...TurboMass Software Installation 176 ...
Страница 179: ...Index ...
Страница 180: ......
Страница 183: ......