CC-LINK Network Communication
WUME-SCHG1C-4
4-66
7.
To check whether the reference waveform has been registered after registering it, send the
“Reference waveform registration status” command (READ). (Command code: 0x0363)
If the response is normal, the following response data will be returned.
Read data
High
Low
Bit15 Bit14 Bit13 Bit12 Bit11 Bit10
Bit9 Bit8 Bit7 Bit6 Bit5 Bit4 Bit3 Bit2 Bit1 Bit0
Reserved
Status
Details of “Status”
0
The "Reference waveform registration execution" command (WRITE) that was executed last time is
successful.
1
The "Reference waveform registration execution" command (WRITE) that was executed last time is
unsuccessful and the reference waveform status has not been updated from the previous state.
(If the power is turned OFF and then ON, the history of beam axis adjustment failures will be erased.)
2
Reference waveform registration is in progress.
8.
To quit beam axis adjustment mode without registering the reference waveform, send the
“Beam axis adjustment end” command (WRITE).(Command code: 0x0364)
Содержание SC-HG1-C
Страница 2: ...WUME SCHG1C 4 2 MEMO ...
Страница 8: ...WUME SCHG1C 4 8 MEMO ...
Страница 16: ...WUME SCHG1C 4 1 8 MEMO ...
Страница 17: ...2 1 2 System Configuration 2 1 Communication unit system 2 2 2 2 Unit types 2 4 ...
Страница 22: ...WUME SCHG1C 4 2 6 MEMO ...
Страница 106: ...WUME SCHG1C 4 4 72 MEMO ...
Страница 107: ...5 1 5 Specifications 5 1 Specifications 5 2 5 2 Dimensions 5 3 ...
Страница 110: ...WUME SCHG1C 4 5 4 MEMO ...
Страница 111: ...6 1 6 Warranty 6 1 When you purchase and use the product 6 2 ...
Страница 114: ...WUME SCHG1C 4 6 4 MEMO ...
Страница 115: ...7 1 7 Maintenance 7 1 Maintenance and Inspection 7 2 7 1 1 Maintenance cautions 7 2 7 1 2 Main inspection items 7 2 ...
Страница 124: ...WUME SCHG1C 4 8 8 MEMO ...
Страница 125: ...Revision History 1st edition 1 27 2017 2nd edition 9 8 2017 3rd edition 9 30 2019 4th edition 12 1 2019 ...
Страница 126: ... MEMO ...
Страница 127: ... MEMO ...