FUNCTION SETTINGS
© Panasonic Industrial Devices SUNX Co., Ltd. 2019
5-24
5-2-11 Response Time (
)
You can set the time until the judgment value (JUDGE.V) is finalized (response time) when the
sensor head starts measurement.
The response time setting determines the average count and average time of measurement as
shown below.
Response time (ms)
Average count
(repetitions)
Average time (ms)
Internal value update cycle
(ms)
3
1
1
1
5
3
3
1
10
8
8
1
100
96
96
4
500
480
480
16
1,000
960
960
32
●
When the Average count is set to 8 times
Sampling cycle (1ms)
Average measurement time (8ms)
Average over 8 times from 1st time and judge
Average over 8 times from 2nd time and judge
Average over 8 times from 3rd time and judge
Internal processing time
Internal processing time(1ms)
<Reference>
●
To respond to instantaneous changes, set a shorter response time. To disregard instantaneous
changes, set a longer response time.
● When the reset input is ON, the indeterminate value display shows “
” until the average
count is reached.
● If the response time is "3" (3ms), the indeterminate value display will show “
” for 1ms.
Содержание HG-S Series
Страница 2: ... Panasonic Industrial Devices SUNX Co Ltd 2019 2 MEMO ...
Страница 7: ...1 1 Chapter 1 INTRODUCTION 1 1 Safety Cautions 1 2 1 2 Safety Information 1 2 ...
Страница 44: ... Panasonic Industrial Devices SUNX Co Ltd 2019 3 24 MEMO ...
Страница 72: ... Panasonic Industrial Devices SUNX Co Ltd 2019 4 28 MEMO ...
Страница 75: ... Panasonic Industrial Devices SUNX Co Ltd 2019 5 3 MEMO ...
Страница 174: ... Panasonic Industrial Devices SUNX Co Ltd 2019 5 102 MEMO ...
Страница 175: ...6 1 Chapter 6 SPECIFICATIONS 6 1 Specifications 6 2 6 2 Dimensions 6 7 ...
Страница 188: ... Panasonic Industrial Devices SUNX Co Ltd 2019 6 14 MEMO ...
Страница 198: ... MEMO ...