FUNCTION SETTINGS
© Panasonic Industrial Devices SUNX Co., Ltd. 2019
5-95
<Reference>
●
In the following cases, calibration cannot be executed. "ERROR" appears in the digital display /
MAIN (white).
•
No magnitude correlation between the target value and detected value (the signs of the
target value and detected value do not match).
•
The adjustment factor that resulted from span adjustment was over twice the factory de-
fault span.
•
The measured value is indeterminate ("------" appears).
•
The measured value is outside the measurement range ("+OVER" or "-OVER" appears).
Setting method
Long-press for 2 seconds
: Select
: Finalize
CALC: Only appears
on master unit
Setting item
Set values
Default value
Calibration selection (CAL.SEL)
Default (DEFALT)
User settings (USER)
DEFAUL