FUNCTION SETTINGS
© Panasonic Industrial Devices SUNX Co., Ltd. 2019
5-94
5-7 Calibration Settings (
)
●
Structure of calibration setting menu
[Calibration selection]
Long-press for 2 seconds
CALC: Only appears on master unit
5-7-1 Calibration Selection (
)
You can perform zero-point adjustment and span adjustment when a sensor head is replaced to
reduce installation error.
Calibration settings
Set values
Function
Calibration selection
(CAL.SEL)
Default (DEFAUL)
Set to the factory default state.
User settings (USER)
Execute calibration using the preset values be-
low, and correct the zero point and span.
Acquired value 1 (CL.SET1)
Move the spindle to the zero point for calibration
and acquire the measured value.
Acquired value 2 (CL.SET2)
Move the spindle to the target point for span ad-
justment and acquire the measured value.
Adjustment value 2 (AJ.VAL2)
Enter a number for the target value.
Zero point
Target point
Span
Step 1
Measure zero point.
Step 2
Measure target point.
Step 3
Enter span target value.
Correct deviation between detected
value and target value.
Содержание HG-S Series
Страница 2: ... Panasonic Industrial Devices SUNX Co Ltd 2019 2 MEMO ...
Страница 7: ...1 1 Chapter 1 INTRODUCTION 1 1 Safety Cautions 1 2 1 2 Safety Information 1 2 ...
Страница 44: ... Panasonic Industrial Devices SUNX Co Ltd 2019 3 24 MEMO ...
Страница 72: ... Panasonic Industrial Devices SUNX Co Ltd 2019 4 28 MEMO ...
Страница 75: ... Panasonic Industrial Devices SUNX Co Ltd 2019 5 3 MEMO ...
Страница 174: ... Panasonic Industrial Devices SUNX Co Ltd 2019 5 102 MEMO ...
Страница 175: ...6 1 Chapter 6 SPECIFICATIONS 6 1 Specifications 6 2 6 2 Dimensions 6 7 ...
Страница 188: ... Panasonic Industrial Devices SUNX Co Ltd 2019 6 14 MEMO ...
Страница 198: ... MEMO ...