Select the 4mA side or the 20mA side for adjustment from the window that appears, as shown in Fig.
140. This description assumes that 4mA will be adjusted first, followed by 20mA.
Input the value reading currently output on the meter and click the "OK" button.
When a value reading is sent, the converter will make adjustments so that the output automatically
becomes 4mA. Verify that the connected meter indicates 1V.
If another adjustment is necessary, input the value reading of the meter into this window again, and
click the "OK" button.
⑩
To adjust 20mA, click the 20mA side of the window shown in Fig. 140, and adjust until the connected
meter indicates 5V, in the same was as with the 4mA adjustment.
⑪
Click the "Cancel" button to exit analog output value adjustment. Click the "OK" button when the
"Adjustment is finished" message box appears as shown in Fig. 138.
⑫
Click the "Cancel" button to halt adjustment in progress, and follow the instructions that appear.
3.7.4.2 Analog output 2 adjustment (Trim Analog 2)
This adjusts the output value of the flowmeter converter
’
s analog output 2. Analog output 2 can be
adjusted by following the same method as described in section 3.7.4.1 "Trim Analog 1", steps
①
to
⑫
.
3.7.5 Cumulative total display and control (Counter/Totalizer cntrl)
Use this for functions such as displaying the cumulative totals, and starting, stopping, and resetting the
count.
3.7.5.1 Cumulative total 1 display and control (Counter/Totalizer cntrl 1)
①
Click the "Diag/Service (T)" menu, and then select and click "Counter/Totalizer cntrl 1" from the
"Counter/Totalizer cntrl" drop-down list.
②
The following window will be displayed (Fig. 141).
Fig.140
Fig.141
Click here and select
which to adjust.